General Specifications | ||
Maximum voltage applied between any terminal and earth ground or between any two terminals | 30 V | |
Operating temperature | -10°C to 50°C | |
Storage temperature | -30°C to 60°C | |
Operating altitude | 2,000 m | |
Storage altitude | 12,000 m | |
Relative humidity (% RH operating without condensation) | Non condensing 90% (10°C to 30°C) 75% (30°C to 40°C) 45% (40°C to 50°C) (Without condensation) | |
Vibration requirements | MIL-T-28800E, Class 2 | |
Drop text requirements | 1 m | |
IP Rating | IEC 60529: IP52 | |
Electromagnetic environment | IEC 61326-1, Portable | |
Safety | IEC 61010-1, Max 30 V to earth, Pollution Degree 2 | |
Power supply | 4 AA NEDA 1.5A IEC LR6 batteries | |
Size (H x W x L) | 52.5 x 84 x 188.5 mm | |
Weight | 515 g | |
DC mA Measurement | ||
Resolution | Range | Accuracy (% of reading + counts) |
0-24 mA | 0.001 mA | 0.010% + 2 μA |
Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) | |
Millivolt Measurement and Source | ||
Resolution | Range | Accuracy (% of reading +counts) |
-10 mV to 75 mV | 0.01 mV | 0.015% + 10 μA |
Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) |
Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus, luctus nec ullamcorper mattis, pulvinar dapibus leo.
Fluke 714B is a handheld, battery-operated instrument that measures and sources a variety of thermocouple types and millivolts. It can calibrate a linear thermocouple transmitter with the mV source function and also measure mA while sourcing temperature. It also provides linear ramp and 25% step auto ramp functionality based on 0% and 100% settings for performing quick, linearity checks.
Designed for calibration professionals
With the intelligent shutdown feature the Fluke 714B remembers settings when powered down so you can quickly replicate tests after powering back up. The calibrator also has one-year and two-year specifications, along with a traceable certificate of calibration. Combine that with its rugged construction, a backlit display, and easy to use interface and you have an ideal test tool for temperature calibration professionals who want a highly accurate, single–function thermocouple temperature calibrator.
General Specifications | ||
Maximum voltage applied between any terminal and earth ground or between any two terminals | 30 V | |
Operating temperature | -10°C to 50°C | |
Storage temperature | -30°C to 60°C | |
Operating altitude | 2,000 m | |
Storage altitude | 12,000 m | |
Relative humidity (% RH operating without condensation) | Non condensing 90% (10°C to 30°C) 75% (30°C to 40°C) 45% (40°C to 50°C) (Without condensation) | |
Vibration requirements | MIL-T-28800E, Class 2 | |
Drop text requirements | 1 m | |
IP Rating | IEC 60529: IP52 | |
Electromagnetic environment | IEC 61326-1, Portable | |
Safety | IEC 61010-1, Max 30 V to earth, Pollution Degree 2 | |
Power supply | 4 AA NEDA 1.5A IEC LR6 batteries | |
Size (H x W x L) | 52.5 x 84 x 188.5 mm | |
Weight | 515 g | |
DC mA Measurement | ||
Resolution | Range | Accuracy (% of reading + counts) |
0-24 mA | 0.001 mA | 0.010% + 2 μA |
Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) | |
Millivolt Measurement and Source | ||
Resolution | Range | Accuracy (% of reading +counts) |
-10 mV to 75 mV | 0.01 mV | 0.015% + 10 μA |
Temperature coefficient | ±(0.002% of reading + 0.002% of range) /°C (< 18°C or > 28°C) |